Passion transducers are designed for reliability and durability. By following recommended care and handling procedures, you can help maximize your transducer's performance and product life. This website provides a summary; please consult the basic system manual for further information.

Multi Beam Technology

Integrated Componded Imaging

Disinfecting the probes

After each use, please disinfect the probes. Ultrasound probes can be disinfected using liquid chemical germicides. The level of disinfection is directly related to the duration of contact with the germicide. Increased contact time produces a higher level of disinfection.
In order for liquid chemical germicides to be effective, all visible residue must be removed during the cleaning process. Thoroughly clean the probe, as described earlier before attempting disinfection.

General Care

Do not soak or saturate transducers with solutions containing alcohol, bleach, ammonium chloride compounds or hydrogen peroxide.

Avoid contact with solutions or coupling gels containing mineral oil or lanolin.

Avoid temperatures above 60°C. Under no circumstances should the transducer be subjected to heat sterilization method. Exposure to temperatures above 60º C will cause permanent damage to the transducer.

Inspect the probe prior to use for damage or degeneration to the housing, strain relief, lens and seal. Do not use a damaged or defective probe.

For More Information please use Ultrasound Passion User Manual.